FLIR IM75-2

74,269.00 GST Extra

Qty:
Compare

Description

Insulation Resistance
4.000 MĪ© Ā± (1.5% + 5 digits) 40.00 MĪ© Ā± (1.5% + 5 digits) 400.0 MĪ© Ā± (3.0% + 5 digits) 4000 MĪ© Ā± (3.0% + 5 digits) 4.1 G to 20.0 GĪ© Ā± (10.0% + 3 digits)
Insulation Test Voltages
50 V for max. 50.0 MĪ©
100 V for max 100.0 MĪ©
250 V for max. 250.0 MĪ©
500 V for max. 500.0 MĪ©
1000 V for max.20 GĪ©
Category Rating
CAT IV-600 V, CAT III-1000 V
MEASUREMENT & ANALYSIS
AC Voltage
1000.0 V: 50 Hz to 60 Hz Ā± (1.5% + 5 digits),
61 Hz to 500 Hz Ā± (2.0% + 5 digits)
Capacitance
1000 ĀµF Ā± (1.2% + 2 digits)
10.00 mF Ā± (1.2% + 20 digits)
Category Rating
CAT IV-600 V, CAT III-1000 V
Continuity
400.0 Ī© Ā± (0.5% + 2 digits)
DC Voltage
1000.0 V Ā± (0.1% + 5 digits)
Earth Bond Resistance
40.00 Ī© Ā± (1.5% + 5 digits)
400.0 Ī© Ā± (1.5% + 3 digits)
4000 Ī© Ā± (1.5% + 3 digits)
40.00 kĪ© Ā± (1.5% + 3 digits)
Frequency [ACV]
400.0 Hz Ā± 5 digits
4.000 kHz Ā± 5 digits
40.00 kHz Ā± 5 digits
Insulation Resistance
4.000 MĪ© Ā± (1.5% + 5 digits) 40.00 MĪ© Ā± (1.5% + 5 digits) 400.0 MĪ© Ā± (3.0% + 5 digits) 4000 MĪ© Ā± (3.0% + 5 digits) 4.1 G to 20.0 GĪ© Ā± (10.0% + 3 digits)
Insulation Test Voltages
50 V for max. 50.0 MĪ©
100 V for max 100.0 MĪ©
250 V for max. 250.0 MĪ©
500 V for max. 500.0 MĪ©
1000 V for max.20 GĪ©
VFD AC Voltage
1000.0 V: 50 Hz to 60 Hz Ā± (1.5% + 5 digits),
61 Hz to 400 Hz, Ā± (5.0% + 5 digits)
COMMUNICATION & DATA STORAGE
Bluetooth range max
10 m (32 ft)
Memory
99 storage locations
POWER
Battery type
6 Ɨ 1.5 V AAA alkaline batteries
ENVIRONMENTAL & CERTIFICATIONS
Certifications
FCC Class B, CE, UL, UKCA, RCM
Drop test
3 m (9.8 ft)
IP Rating
IP54 dust-protected and splash-resistant
Operating Temperature
0Ā°C to 50Ā°C (32Ā°F to 122Ā°F)
Warranty
https://www.flir.com/testwarranty